一、产品简介:

水泥雷氏沸煮箱是水泥净浆体积的安定性**检测仪器,水泥雷氏沸煮箱是根据JC/T955-2005标准《水泥标准稠度用水量、凝结时间,安全性检验方法》设计制作,能自动控制箱体内水升温至沸腾和保持沸腾的时间,以检定水泥净浆体积的安定性(即雷氏法和试饼法),对升温保温均能自动控制,也可人工控制。

Product introduction of cement boiling box:

Cement Lei Boiler is a stability testing instrument for the volume of cement paste. Cement Lei Boiler is designed and manufactured according to JC / T955-2005 Standard "Concrete Standard Consistency Water Consumption, Setting Time, Safety Inspection Method" It can automatically control the temperature of the water in the box to boil and keep boiling time, in order to verify the stability of the cement slurry volume (ie, the Lei's method and the test cake method), can automatically control the heating and insulation, and can also be controlled manually.


二、技术参数:

★A高沸煮温度  A℃    

★容积  31L    

★升温时间(20℃升至A℃)  30±5min    

★横沸时间  3h±5min    

★管状加热器功率  共两组(1kW和3kW)  

★包装尺寸  82×40×53cm  

★毛重/净重  22/35kg  

★电压功率  220V/4KW  

Technical parameters of cement boiling box:

★ A high boiling temperature A ℃

★ Volume 31L

★ Heating time (20 ℃ to A ℃) 30 ± 5min

★ Horizontal boiling time 3h ± 5min

★ Tubular heater power two sets (1kW and 3kW)

★ Packing size 82 × 40 × 53cm

Gross weight / net weight 22 / 35kg

★ Voltage power 220V / 4KW


三、产品结构:

由沸煮箱体,控制器和连接电缆组成。

沸煮箱体主要由箱盖1,内外箱壁2,蓖板3,保温层4,电加热管5,放水阀6,把手7,出气孔8,水封槽9和接线盒10组成。

箱体和蓖板均用A锈钢制成,外表面磨花。控制器由机箱、控制器逻辑板FZH-3、底板和面板组成。

底板上装有交流继电器、逻辑板插座,控制逻辑板**插在该插座上,稳压电源也装在逻辑板上。面板上装有多种控制开关、按钮和指示灯。壳体后面有电源进线插座,负载线引出插座和保险。面板与底板用十六脚插头座连接。便于维修。

Product structure of cement boiling box:

It consists of a boiling cabinet, a controller and a connecting cable.

The boiling box body is mainly composed of a box cover 1, inner and outer box walls 2, castor 3, heat insulation layer 4, electric heating pipe 5, water discharge valve 6, handle 7, air outlet 8, water seal groove 9 and junction box 10.

The cabinet and cast iron are made of A rust steel, and the outer surface is ground. The controller consists of a chassis, a controller logic board FZH-3, a backplane, and a panel.

An AC relay and a logic board socket are installed on the bottom plate. The control logic board ** is inserted into this socket, and the regulated power supply is also mounted on the logic board. Various control switches, buttons and indicators are installed on the panel. There is a power inlet socket on the back of the housing, a load cable leads out of the socket and a fuse. The panel and the bottom plate are connected by a sixteen-pin plug socket. Easy maintenance.


四、使用方法:

★合上电源闸刀,打开电源开关,电源指示灯亮。

★“方式选择”开关拨到“自动”位置。

★按“启动”按钮,此时“加热”入“沸煮”指示灯同时点亮,沸煮箱将根据预定程序计时开始工作,这时计时显示灯以次/2秒的速度闪亮。加温30分钟→“加热”指示灯灭“沸煮”三小时→“沸煮”指示灯灭→蜂鸣器发出叫声。

★试验结束,关闭电源开关,拉开电源闸刀,取出试样。

Fourth, the use of cement boiling box

★ Close the power switch, turn on the power switch, and the power indicator light is on.

★ Turn the "Mode Selection" switch to the "Auto" position.

★ Press the "Start" button, the "Heating" and "Boiling" indicator lights at the same time, and the boiling box will start to work according to the predetermined program. At this time, the timer display lamp will flash at the rate of 2 seconds. Heating for 30 minutes → the “heating” indicator goes off for “three hours” → the “boiling” indicator goes off → the buzzer sounds.

★ After the test, turn off the power switch, pull open the power switch blade, and remove the sample.


五、操作注意事项:

★注意:※※箱内无水严禁通电!!!

★使用沸煮箱时,要认真阅读技术说明书,熟悉技术指标、工作性能、使用方法、注意事项,严格遵照仪器使用说明书的规定步骤进行操作。

★初次使用雷氏沸煮箱人员,必须在熟练人员指导下进行操作,熟练掌握后方可进行独立操作。

★实验时使用的雷氏沸煮箱,要布局合理,摆放整齐,便于操作,观察及记录等

Fifth, precautions for operation of cement boiling box:

★ Note: ※ ※ There is no water in the box. !! !!

★ When using the boiling box, read the technical manual carefully, be familiar with the technical indicators, working performance, use methods, and precautions, and strictly follow the steps specified in the instrument's manual.

★ For the first time using the Lei Boiling Boiling Box, you must operate it under the guidance of skilled personnel, and you can perform independent operation only after mastering it.

★ The Lei's boiler used in the experiment should be reasonably arranged and neatly arranged for easy operation, observation and recording.


五、操作规程:

★根据用户要求试饼法和雷氏法同时使用,以对比水泥安定试验结果,特将篦板高度降低,为此请用户使用篦板时(包括只做雷氏夹法)务必置于试饼架之上。

★沸煮箱内充水180mm(以内箱体底部起算)将经过养护的试饼或雷氏夹由玻璃板上取下来平放在试饼架或者雷氏夹两指针朝上,试模横放于篦板上。 

★接通电气控制箱电源,启动“自动”开关,沸煮箱内的水于30分钟左右后沸腾,一组3kW电热器自动停止工作,指示灯灭。再煮3h沸煮箱内另一组1kW电热器自动停止工作,指示灯灭。此时数字显示为210 min,电气控制箱内蜂鸣器发出声响,表示工作结束。煮毕将水由电热水咀放出,打开箱盖待箱体冷却至室温,取出试件进行检测。 

★如沸煮箱内充水温度低于20℃时,可先将电气控制箱上手动自动切换开关放在手动位置、升温恒温开关放在升温位置。将水升温至20℃左右,将手动、自动开关切换到自动位置即可自动运行。 

★沸煮箱内必须用洁净淡水,本设备久用后,箱内可能累积水垢,用户可酌情定期清洗

Fifth, the operating procedures of the cement boiling box:

★ According to user requirements, the test cake method and Rayleigh method are used at the same time to compare the results of the cement stability test. The height of the concrete plate is reduced. For this reason, please use the concrete plate (including the Rayleigh clamp method) to place the test cake. Above the shelf.

★ Fill the boiling box with 180mm of water (from the bottom of the inner box). Remove the cured test cake or Lei's clamp from the glass plate and place it on the test cake rack or the Lei's clamp. On the sampan. Zh

★ Turn on the power of the electric control box and start the "auto" switch. The water in the boiling box will boil in about 30 minutes. A group of 3kW heaters will automatically stop working and the indicator light will go out. After another 3 hours, another set of 1kW electric heaters in the boiling box will automatically stop working, and the indicator light will go out. At this time, the digital display is 210 minutes, and the buzzer in the electrical control box sounds, indicating that the work is over. After cooking, release the water from the electric hot water nozzle, open the lid of the box and allow the box to cool to room temperature. Take out the test piece for testing. Zh

★ If the filling temperature in the boiling box is lower than 20 ℃, you can first place the manual automatic switch on the electric control box in the manual position and the temperature-rising thermostatic switch in the temperature-rising position. Warm the water to about 20 ° C and switch the manual and automatic switches to the automatic position to run automatically. Zh

★ Clean fresh water must be used in the boiling box. After the device is used for a long time, scale may accumulate in the box. Users can clean it regularly as appropriate.


六、开机前的准备:

★用包装箱内带的2.5平方毫米的三芯护套电缆连接控制器电源插座。

★用专用电缆连接控制器和箱体,连接方法见图1(出厂前电缆与箱体已连接好)。

★控制器和沸煮箱的接地螺栓牢靠接地(A能接电线零线)。

★检查供电电源:允许范围为~220V+10%,50HZ。

★接好电源电缆:建议使用20A的单相闸刀作为总电源开关,电源电缆的绿色线接相线,红色线接零线。

★按照要求在沸煮箱内加水(180毫米),放入试样。

★在水封槽内加水,盖上箱盖。

★检查控制器面板开关位置:“电源”开关置于“关”;“方式选择”开关置于“自动”位置。

Fifth, the operating procedures of the cement boiling box:

★ According to user requirements, the test cake method and Rayleigh method are used at the same time to compare the results of the cement stability test. The height of the concrete plate is reduced. For this reason, please use the concrete plate (including the Rayleigh clamp method) to place the test cake. Above the shelf.

★ Fill the boiling box with 180mm of water (from the bottom of the inner box). Remove the cured test cake or Lei's clamp from the glass plate and place it on the test cake rack or the Lei's clamp. On the sampan. Zh

★ Turn on the power of the electric control box and start the "auto" switch. The water in the boiling box will boil in about 30 minutes. A group of 3kW heaters will automatically stop working and the indicator light will go out. After another 3 hours, another set of 1kW electric heaters in the boiling box will automatically stop working, and the indicator light will go out. At this time, the digital display is 210 minutes, and the buzzer in the electrical control box sounds, indicating that the work is over. After cooking, release the water from the electric hot water nozzle, open the lid of the box and allow the box to cool to room temperature. Take out the test piece for testing. Zh

★ If the filling temperature in the boiling box is lower than 20 ℃, you can first place the manual automatic switch on the electric control box in the manual position and the temperature-rising thermostatic switch in the temperature-rising position. Warm the water to about 20 ° C and switch the manual and automatic switches to the automatic position to run automatically. Zh

★ Clean fresh water must be used in the boiling box. After the device is used for a long time, scale may accumulate in the box. Users can clean it regularly as appropriate.


水泥安定性沸煮箱
水泥雷氏沸煮箱是水泥净浆体积的安定性**检测仪器,水泥雷氏沸煮箱是根据JC/T955-2005标准《水泥标准稠度用水量、凝结时间,安全性检验方法》设计制作,能自动控制箱体内水升温至沸腾和保持沸腾的时间,以检定水泥净浆体积的安定性(即雷氏法和试饼法),对升温保温均能自动控制,也可人工控制。
长按识别二维码查看详情
长按图片保存/分享

水泥安定性沸煮箱

询盘

咨询表单:

  • 请输入验证码

咨询内容:


你还没有添加任何产品

加入成功

扫一扫关注我们

Copyright @ 2018 上海荣计达仪器科技有限公司.All Rights Reserved   沪ICP备11009963号-4     

上海市闵行区浦江镇沈杜公路3259号                                                             网站地图

Copyright @ VillaGrandis All Rights Reserved某某某有限公司版权所有        

ICP备案号180006020号-3

 

添加微信好友,详细了解产品
使用企业微信
“扫一扫”加入群聊
复制成功
添加微信好友,详细了解产品
我知道了